Measurement of SQUID noise levels for SuperCDMS SNOLAB detectors - Final Paper
Lee, Maxwell.
SuperCDMS SNOLAB is a second generation direct dark matter search. In the SuperCDMS SNOLAB experiment, detectors are able to pick up from signals from dark matter nuclear recoil interactions which occur inside the bulk of the detectors. These interactions produce both phonon and charge signals. HEMTs read out charge signals whereas TES are used to detect phonon signals which are then read out by SQUID amplifiers. SQUID amplifiers must add negligible noise to the TES intrinsic noise which has bee... Show Full Abstract
SuperCDMS SNOLAB is a second generation direct dark matter search. In the SuperCDMS SNOLAB experiment, detectors are able to pick up from signals from dark matter nuclear recoil interactions which occur inside the bulk of the detectors. These interactions produce both phonon and charge signals. HEMTs read out charge signals whereas TES are used to detect phonon signals which are then read out by SQUID amplifiers. SQUID amplifiers must add negligible noise to the TES intrinsic noise which has been previously measured and is approximately 50pA/(root)Hz down to 100Hz for ease of signal distinguishability in dark matter nuclear interactions. The intrinsic noise level of the SQUID was tested in the SLAC 300mK fridge and determined to provide adequately low levels of noise with a floor of approximately 3pA/(root)Hz. Furthermore, a 10x amplifier was tested for addition of extraneous noise. This noise was investigated with and without this amplifier, and it was found that it did not add a significant amount of noise to the intrinsic SQUID noise. Show Partial Abstract
  • Interest Categories: Other